The Animal Disease Rapid Diagnostic Instrument is a brand-new testing device recently launched by our company. This system employs the enzyme-linked immunosorbent assay (ELISA) method in conjunction with corresponding reagents to enable qualitative or quantitative rapid diagnosis of animal diseases. For example: Porcine reproductive and respiratory syndrome virus (PRRSV), classical swine fever virus (CSFV), pseudorabies virus (PRV), PRV gE protein, porcine circovirus (PCV), Japanese encephalitis virus (JEV), foot-and-mouth disease virus (FMDV) 3ABC protein, FMDV IgG, porcine parvovirus (PPV), avian influenza, bluetongue disease, white spot syndrome, bovine contagious pleuropneumonia, rinderpest, sheep pox, and goat pox, among others. Customized testing projects can also be developed according to customer requirements.
10.4-inch color LCD touchscreen input with a large display for quick and easy operation.
Board layout and save/import functions significantly improve testing efficiency for frequently used projects.
Full-screen display of the 96-well visualization layout operation interface.
Intelligent light source switch to extend its lifespan.
Automatic storage of standard curves, with the option for users to adjust standard curves.
Built-in embedded system, enabling operation, storage, and printing without an external computer.
Template functionality to standardize testing workflows and reduce operator workload.
Automatic zeroing available in both single-wavelength and dual-wavelength testing modes.
Massive storage capacity, capable of storing at least 1 million plates (96) of test data and 500 programmed projects, with support for data backup, recovery, and merging.
New Windows-based color graphical user interface with a fully Chinese-language operating system, supporting mouse, keyboard, and multiple input methods.
Comprehensive quality control functions, including Westguard multi-rule quality control, instant method quality control, and various quality control charts and parameter calculations, with quality control alarm functionality.
Can perform 24 different tests on the same plate, with distinguishable displays, and can test both qualitative and quantitative projects on the same plate.
Intelligent calculation of test results to ensure accuracy and validity, with automatic alarm functionality.
Offers multiple printing options and various formats of Chinese reports, with customizable report templates.
Optional external PC software support for network connectivity, aligning with digital development trends, with customizable upload capabilities.
Dedicated system maintenance menu automatically assesses the instrument's current status and performance metrics, with the ability to print relevant data.
Supports various calculation methods, including absorbance, Cut-Off qualitative calculation, single-point calibration, linear regression, exponential regression, logarithmic regression, double logarithmic regression, log-logit, power regression, four-parameter regression, and enzyme inhibition rate.
Wavelength range | 330-1100 nm |
Filter configuration | 10 filter positions, standard configuration includes 405 nm, 450 nm, 492 nm, 630 nm, optional 6 custom wavelengths |
Absorbance Range | 0.000–4.000 A |
Number of Optical Channels | 8-channel optical path detection, plus one independent reference channel |
Display Stability | ≤±0.002A or ≤0.5%T/10min |
Display Error (Accuracy) | ≤±0.005A or ±0.2%T |
Repeatability | ≤0.2% or ≤0.2%T |
Sensitivity | ≥0.01 (L/mg) |
Channel difference | ≤0.01A |
Wavelength indication error | ±1 nm |
Half-width | ≤8 nm |
Peak transmittance ratio | ≥35% |
Resolution | 0.001 A (display), 0.0001A (internal calculation) |
Reading speed | Single wavelength ≤3 seconds/96 wells, Dual wavelength ≤6 seconds/96 wells |
Shaking function | Adjustable speed and time |
Light source type | Halogen tungsten lamp |
Plate type | Standard 96-well or other types of enzyme-linked immunosorbent assay (ELISA) plates, strips |
Suitable well types | Flat-bottom, U-shaped, and V-shaped |
Display method | 10.4-inch color LCD display |
Input method | Touchscreen input, optional mouse and keyboard |
Printing | Built-in thermal printer, compatible with external printers |
Interfaces | DVI port, VGA port, Ethernet port, USB port, audio jack, RS232 serial port |
Operating environment | Temperature 5-40°C; Humidity 15%-80% |
Power voltage | 220V±10%, 50/60Hz |
Dimensions and weight | 470mm×350mm×210mm (L×W×H), 11.5Kg |