Sulfide blowing apparatus

Sulfide blowing apparatus

  • Product Number:LD-LHW60
  • Product Difference:Simultaneously process six samples; sample pretreatment is required for water quality sulfide determination

Product Introduction

The principle of the sulfide blow-through analyzer is as follows: sulfides in water samples are acidified, and the resulting hydrogen sulfide is carried by the carrier gas (nitrogen) into the absorption bottle/colorimetric tube, where it is absorbed by the absorption solution (zinc acetate-sodium acetate solution or 2% sodium hydroxide solution). The absorbed sulfide ions in the absorption bottle/colorimetric tube are then analyzed and measured using the appropriate analytical method.

The sulfur compound blowing instrument is developed and manufactured in accordance with the new national standard for the determination of sulfur compounds in water quality—the methylene blue spectrophotometric method (titrimetric method)—established in 2022. It is required for the sample pretreatment in the determination of sulfur compounds in water quality.

Product Features

  • The hydrochloric acid inlet, gas inlet, and outlet are integrated into a single unit yet remain independent, ensuring that operations do not interfere with one another.

  • Constant-temperature water bath heating with a rapid automatic temperature control system, allowing the heating temperature to be manually set.

  • The sample rack and gas distribution system are adjusted via a height-adjustable knob, allowing for easy installation of samples on the front and convenient operation.

  • The nitrogen flow rate for each sample can be independently controlled, adjusted, or shut off, with a needle valve gas flow meter accurately controlling and displaying the total gas consumption.

  • Large-capacity reaction flasks enable simultaneous processing of multiple samples.

Product parameters

Sample count

6 positions
Heating methodAutomatic temperature-controlled constant-temperature water bath

Heating power

1000W, uniform heating
Nitrogen flow meter control0-3 L/min
Temperature rangeRoom temperature to 99.9°C
Display methodDigital display
Temperature control accuracy±1°C

Nitrogen inlet pressure

0.1 MPa
Nitrogen flow branch0-0.6 L/min
Operating power supplyAC (220±22) V, 50 Hz
Ambient temperature(5-35)°C
Ambient humidity(0-95)% RH