The environmental radon detector is suitable for measuring radon volume activity in ambient air, soil, and water, as well as surface radon emission rates in soil and building materials. It can be used for environmental monitoring, geological prospecting, radiation protection, nuclear accident monitoring, radiation dose assessment, earthquake prediction, and educational purposes.
Compact, lightweight, and portable
Dimensions: 330mm × 210mm × 170mm, Weight: 2.5KG.
Advanced, Accurate, Reliable
The CD2 Smart Environmental Radon Monitor employs an electrostatic collection-semiconductor sensor-alpha energy spectrum analysis method for radon measurement. The pump draws the sample gas (air) into the electrostatic collection chamber. Within this chamber, a high-voltage electric field adsorbs RaA (²¹⁸Po), the first-generation decay product of ²²²Rn, onto the surface of the semiconductor alpha-ray sensor (cathode). Through energy spectrum analysis, the alpha particle count rate of RaA is measured to quantitatively determine the volume activity of ²²²Rn.
Utilizing the energy spectrum measurement of alpha particles from the short-lived daughter product of 222Rn (218Po with a half-life of 3 minutes) may effectively mitigate thorium radon interference during soil radon measurements. Additionally, the short half-life of the measured daughter product enables the detector to return to low background levels within a brief timeframe during high-activity measurements (e.g., soil radon).
Built-in climate sensors precisely measure gas temperature, humidity, and atmospheric pressure within the static chamber, indicating desiccator status and enabling gas volume correction and temperature-adsorption rate correction.
Intelligent and User-Friendly
The CD2 Smart Environmental Radon Monitor utilizes an intelligent touchscreen platform based on an ARM processor and Android 4.4 system for data acquisition, processing, display, and printing. This enables the PRn700 series smart environmental radon monitors to offer multiple intuitive and user-friendly human-machine interaction modes, including images, sounds, wired/wireless networks, and touch sensing.
The computing platform built on the ARM processor and Android 4.4 operating system delivers robust data processing capabilities. It supports diverse data connectivity options including Wi-Fi, Bluetooth, USB (HOSTDEVICE mode), RJ45, and RS232, enabling users to update software.
Software design aligns with mainstream smart touchscreen device operation modes, including intelligent backlighting, automatic shutdown when idle, and confirmation prompts for critical operations. The product is easy to operate, allowing users to master its functions after brief, simple exploration.
Handheld Bluetooth printer with self-adhesive report labels. One-touch printing enables instant data preservation—simply peel and stick. The host unit provides charging for the printer, eliminating the hassle of finding power sources for field printing!
Comprehensive Accessories and Functionality
Equipped with various professional accessories for measuring radon activity in soil, building materials, water, and more. Featuring mature, reliable technical solutions, a highly integrated platform, and a robust software environment, the device boasts a compact structure and enhanced reliability.
| Electrostatic Chamber | Volume 700 ml, high voltage 2500–3000 V |
| Detector | Semiconductor planar silicon detector, effective detector area 572 mm²; Alpha particle energy measurement range: 0–10 MeV, energy resolution: 37 keV (FWHM) |
| Background Count Rate | ≤0.003 cpm |
| Detection Sensitivity | 0.2 cpm /pCi/L |
| Detection Limit | ≤1 Bq/m³ (120 min, 2σ) |
| Measurement Range | Air Radon: (1–999999) Bq/m³ Soil Radon: (300 to 999999) Bq/m³ Radon in water: (0.001 to 100.00) Bq/L Radon desorption rate: (0.001 to 1000.000) Bq/[m²•s] |
| Measurement uncertainty | ≤10% (k=2) |
| Annual Volume Activity Response Offset | ≤±20% |
| Relative Intrinsic Error | ≤±20% |
| Power Supply | Lithium-ion Rechargeable Battery: 11.1V, 5400mAh Charger Input: AC (110–240)V, Output: 12.6V/2A |
| Operating Environment | Temperature (5–40)°C , Humidity ≤90%RH |
| Display | 5.5-inch 5-point capacitive touch LCD screen |
| Gas Sampling Method | Active pump suction, Pump Rate: 2L/min (without vacuum load) |
| Measurement Time (under typical conditions) | Air Radon: 20-120 min, Soil Radon: 9-17 min, Radon Desorption Rate: 60-300 min (excluding gas collection time) |
| Climate Sensor | Temperature: Measurement range (0–50°C), accuracy ±0.5°C; Pressure: Measurement range (300–1100) |
| Dimensions & Weight | 330mm × 210mm × 170mm, 2.5 kg (includes protective case, filter, charger) |